Stage System
Stage Menu Commands
Rev. D
Dimension 3100 Manual
77
5. To exit the
Move To
(
x,y
) dialog box, click the
Quit
button.
5.2.5 Set Reference
The
Set Reference
panel is used to set the origin point on the sample surface to be used for all
subsequent
Move To
and
Programmed Move
operations. It is important to verify the reference
point (0,0) with programmed move sequences, since all moves are relative to the current origin. For
example, a program designed to move the stage to the four corners of a square sample when the
origin (0, 0) is set to the upper-left corner will
not
probe the same positions if the origin is changed
to the lower-right corner. It will avoid confusion to select a standard position as the usual origin
reference (e.g., lower-left corner), then reuse the same reference position with all samples.
On samples having a grid-like aspect (e.g., integrated circuits), the reference may be initially
defined from a line (two points) rather than a single point. The reference line is defined parallel to
some feature on the sample surface (e.g., an electronic trace on an integrated circuit). You may
complete this by defining two points: an origin and a second point. Defining the second point
compensates for sample rotation. To set a reference on the sample, complete the following:
1. Select the
Focus Surface
function under the
Stage
menu to focus on the sample and move
the sample to the desired origin.
2. Select the
Set Reference
option under the
Stage
menu. The screen offers five options to the
CAUTION:
Always verify that the tip is off the surface before attempting stage
movements. If manual stage movements are attempted during
engagement (by turning the leadscrew knobs on the stage’s X-Y
slide assemblies) the tip and/or sample may be damaged.
ATTENTION:
Il est impératif de toujours s’assurer que la pointe ne touche pas la
surface avant de bouger la platine porte-échantillon. Si la platine
porte échantillons est déplacée manuellement (en tournant les vis
des moteurs de la platine) la pointe et/ou l’échantillon pourraient
être endommagés.
ATTENTION:
Überprüfen Sie immer zunächst, daß die Meßspitze nicht mehr auf
der Oberfläche ist, ehe Sie den XY-Verschiebetisch bewegen. Wenn
der XY-Verschiebetisch von Hand bewegt wird (indem die
Drehknöpfe an den Gewindestangen des XY-Verschiebetisches
gedreht werden), während sich das Mikroskop im Engage-Zustand
befindet, können Meßspitze und/oder Probe beschädigt werden.
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......