Calibration
Linearity Correction
320
Dimension 3100 Manual
Rev. D
Fine Adjustment
Initial adjustment is usually adequate; however, if more precision is desired, use the following fine
adjustment techniques to adjust
Fast mag1
.
1. Use the same procedure for adjusting Fast mag0.
2. As before, set the
Zoom
box for the beginning of the scan and then check the ending.
Because the scan is small, use a
Zoom
box up to one-half as large as the scan.
3. If the end of the scan is larger than the beginning, reduce the
Fast mag1
value. If the end is
too small, increase the value of
Fast mag1
.
17.4.3 Adjust Slow mag1
1. With the
Scan size
set to
150V
, select
Microscope
>
Calibrate
>
Scanner
.
2. Select
Slow mag1
and input the value from
Fast mag1
.
Note:
For medium-sized scanners (C to F), further adjustments should not be
necessary. The
Slow mag1
value is usually 100-120 percent of the
Fast mag1
value.
3. If more precision is desired, adjust
Slow mag1
using the same procedure for adjusting
Slow
mag0
.
4. Set the
Zoom
box for the beginning of the scan and then check against the ending. Because
the scan is small, use a
Zoom
box up to one-half as large as the scan.
5. If the end of the scan is larger than the beginning, decrease the
Slow mag1
value. If the end
is too small, increase the value of
Slow mag1
.
6. Wait one complete frame with the new value before readjusting the
Slow mag1
value.
7. Check the final result by capturing an image and checking it with the
Offline
>
Modify
>
Zoom
window. The scanner is now ready for calibration of the X and Y parameters.
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......