Electric Techniques
Electric Force Microscopy
278
Dimension 3100 Manual
Rev. D
Setting Analog 2 in the Software
If you have chosen a configuration where Analog 2 applies to the tip or sample, you must enable
Analog
in the software. The Analog 2 parameter appears in both the
Feedback Controls
and
Interleave Controls
panels.
Note:
For Version 4.23 and lower, Analog 2 appears only in the
Feedback Controls
panel.
Instructions for Enabling Analog 2 in Software
1. Select
Di
>
Microscope Select
>
Edit
>
Advanced
, and set Analog 2 to
User defined
.
2. Click
OK
to exit both dialog boxes.
Note:
For software versions 4.23 and lower, select
Microscope
>
Calibrate
>
Detector
to display the
Detectors Parameters
window. Switch the
Allow in
attenuation
field to
Disallow
.
Instructions for Disabling Analog 2 in Software
1. Select
Di
>
Microscope Select
>
Edit
>
Advanced
, and set Analog 2 to
Atten switch
.
2. Click
OK
to exit both dialog boxes.
Note:
Remember to set
Allow in attenuation
to
Allow
when finished. For all other
configurations it should be left on
Allow
.