Scan Routine Parameters
7-5
Figure 7-5: Scan Parameters Dialog Box: Scan Location
NOTE –
Because the numeric values of the X and Y location are often not known, in most
cases you must determine the stage location by using the
Sample Positioning
window. From
the menu in this window, select
Edit > Define Scan Location/Length
.
Scan Length
Scan lengths from 50 µm to 55,000 µm (50 mm) are possible using the standard Dektak 150 system.
1
Click any parameter in the
Scan Parameters
section of the
Scan Routines
window (see
display the
Nominal Parameters
tab of the
Scan Parameters
dialog box.
2
In the
Length
field, enter the desired scan length using the keyboard (see
).
NOTE –
The scan length is expressed in microns (µm).
3
Click
OK
to close the dialog box and enter the length into the scan program.
Appears only if your
system includes the
optional x-y auto
stage.
Содержание Dektak 150
Страница 1: ......
Страница 2: ...P N 980 294 Standard P N 980 298 Cleanroom Revision A DEKTAK 150 SURFACE PROFILER USER S MANUAL ...
Страница 7: ......
Страница 17: ......
Страница 81: ......
Страница 105: ......
Страница 149: ......
Страница 191: ......
Страница 251: ......
Страница 257: ...A 6 Facilities Specifications Figure A 6 Dektak 150 without Enclosure Side View ...
Страница 258: ...Facilities Specifications A 7 Figure A 7 Dektak 150 Dimensions without Enclosure Top View ...
Страница 259: ...A 8 Facilities Specifications Figure A 8 Dektak 150 without Enclosure Center of Mass ...
Страница 273: ......
Страница 283: ......
Страница 320: ...3D Mapping Function E 37 Figure E 26 Database Options Button Figure E 27 Database Options Dialog Box ...
Страница 321: ......
Страница 331: ......
Страница 333: ......
Страница 336: ......
Страница 337: ......
Страница 349: ......