Pax-i3D Smart(PHT-30LFO) Installation Manual
28
2 Choosing an Installation Site
2.7
Installing the Warning Lamp and Door Interlock Switch
Refer to Appendix
A
for a complete installation guide.
●
This system can be equipped with a warning lamp and the door interlock switch which are
activated when the X-Ray is energized.
●
The warning lamp and the door interlock switch are not included with the equipment.
● The warning lamp and the door interlock switch must be installed by a qualified technician.
2.8
Installing the Emergency Stop Switch
Refer to Appendix
B
for a complete installation guide.
●
Install the emergency stop switch along the main power cable in the central distribution panel.
●
Install this switch so that it is within easy reach of the operator but cannot be accidentally
pressed.
●
The switch must be a fool-proof model.
Содержание PaX-i3D Smart
Страница 1: ...PHT 30LFO HARDWARE and SOFTWARE E n g l i s h Installation Manual Version 1 0 0...
Страница 2: ......
Страница 3: ......
Страница 72: ...This page is intentionally left blank...
Страница 82: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 137: ...10 Technical Specifications...
Страница 139: ...Pax i3D Smart PHT 30LFO Installation Manual 139 10 Technical Specifications Dimension of equipment PANO CT Unit mm...
Страница 140: ...Pax i3D Smart PHT 30LFO Installation Manual 140 10 Technical Specifications PANO CT CEPH Unit mm...
Страница 141: ...Pax i3D Smart PHT 30LFO Installation Manual 141 10 Technical Specifications PANO CT CEPH OneShot type Unit mm...
Страница 145: ...Pax i3D Smart PHT 30LFO Installation Manual 145 Appendix Schematic diagram Components supplied...
Страница 160: ......
Страница 161: ...13 Samsung 1 ro 2 gil Hwaseong si Gyeonggi do Korea Postal Code 445 170...