Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
59 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
3.5. CT sensor: Bright calibration
•
Stay outside of the X-ray shielding room during Bright Calibration data
achievement.
3.5.1. Bright calibration: 2 points
1. Run the C:\VCaptureSW\Acquisition\CBCT\WidePANO\VACAP.exe file.
2. In the
Settings
tab and make sure that the parameters under [BrightCalibration] are as
shown below and click the
Save
button if something is changed.