Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
172 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
5.2. Test Method
1. Complete the collimator set up and acquire the image in the stopped state.
2. Run C:\VCaptureSW\Utility\EzEval_Phantom\EzEval.exe program.
3. Enter 90% as the threshold standard and check if it satisfies the Collimator Alignment standard.
5.3. Test Result
High-Resolution Lateral Mode
Parameter
LSL
Measurement value
USL
Pass/Fail
Threshold:
90%
Upper
20
40
Lower
20
40
Left
1
5
Right
1
5
TOP
LEFT
RIGHT
BOTTOM