2.
PA Mode Start Position Evaluation
2.1. PA Mode starting position standard
Parameter
LSL (Lower Specification Limit)
USL (Upper Specification Limit)
PA Start Position
1160
1180
2.2. Phantom Specification
CEPH Phantom QUART Phantom
2.3. Test Method
1. Acquire images in High-Resolution PA Mode.
2. The evaluation is conducted in the same manner as in the evaluation method in Section 1.3.
2.4. Test Result
Parameter
LSL
Measurement value
USL
Pass/Fail
PA Start Position
1160
1180