Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
142 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
Chapter 3. CEPH Image Optimization
1.
CEPH Sensor Calibration
1.1. Before CEPH sensor calibration
•
VAKCAP.com program is required for CEPH Sensor Calibration.
Path:
C:\VCaptureSW\Acquisition\Ceph\WideCeph\Norm\
folder
•
Before the CEPH Sensor Calibration operation begins, the operator
must backup the existing CEPH Sensor Calibration folder.
•
CEPH Sensor Calibration file folder path:
C:\VCaptureSW\Acquisition\CEPH\WideCeph\Norm\CAL
1. Remove the Normal Chin Block from the equipment.
2. Remove the CEPH ear rods and fold the nasal positioned to avoid CEPH sensor blocking.
3. Disassemble the Tube cases (Refer to Chapter 1. Hardware Part Service Guide).
4. Rename the CAL folder (C:\VCaptureSW\Acquisition\Ceph\WideCeph\Norm\CAL) name for
backup.