Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
137 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
Chapter 5. PANO Image Evaluation
1.
Line Pair Resolution evaluation
1.1. Resolution standards
Parameter
LSL (Lower Specification Limit)
USL (Upper Specification Limit)
Lp/mm
3.1
-
1.2. Phantom Specification
QUART Phantom
QUART - Ball Phantom
1.3. Test Method
1. Run the Console Software and click the PANO button to go into PANO mode.
2. Choose HD Mode/MAN/Normal mode/exposure condition 70 kVp, 7 mA as the image option and
click the Confirm button.
3. Attach PANO BALL P Quart Phantom Ass'y on Chinrest.
4. Attach 1 .8 mm Copper Filter on X-Ray Tube.
5. Click the READY button, and acquire/save the image according to the Console Software
guidance.
6. Open the 2D Viewer and save the image.
7. Make sure that the lines can be seen clearly at 3.1 lp/mm.
1.4. Test Result
Parameter
LSL
Measurement value
USL
Pass/Fail
Lp/mm
3.1
-