CT Number Accuracy evaluation
5.1. CT Number standards
Parameter
LSL (Lower Specification Limit)
USL (Upper Specification Limit)
Air
-1030
-900
Water
-20
20
Teflon
900
1100
5.2. Phantom Specification
CT Number Phantom