Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
101 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
1.3. Test Method
1. Run Console Software and click CBCT to go into CBCT mode.
2. Choose FOV 8x8/Adult/Normal/High Resolution/Standard as the image option and click the
Confirm button.
3. Put the Phantom jig on the unit and make the phantom jig level by using bubble level and three
align pins.
4. Put the DVT Phantom on the phantom jig.
5. Press the
Ready
button and acquire an image.
6. When reconstruction is completed, save the CT folder name in the C:\VCaptureSW\ImageOutput
folder as CT_DVT
7. Run the DVTpro Software and conduct the evaluation.
1.4. Test Result
Parameter
LSL
Measurement value
USL
Pass/Fail
MTF 10% <lp/mm>
1.00
-