![Varian Turbo-V300HT Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/varian/turbo-v300ht/turbo-v300ht_instruction-manual_3692034036.webp)
TECHNICAL INFORMATION
33
87-900-875-01(C)
VENT DEVICE INSTALLATION
The vent device mod. 969-9831 allows to avoid
undesired venting of the pump during a temporary
power failure (adjustable time up to 36 min.), and
enables an automatic vent operation.
Unscrew and remove the threaded plug (see figure
below).
Screw the flange mod. 969-9109 on the pump, tak-
ing care of the o-ring right position.
Assemble the seal ring and lock the vent device in
position using the KF klamp.
VIBRATION ISOLATOR INSTALLATION
Four vibration isolators for ISO and CFF inlet
flange version pumps are available as accessories.
The four model part numbers are the following:
−
model 969-9342 for ISO 100 flange;
−
model 969-9343 for ISO 160 flange;
−
model 969-9332 for CFF 6” flange;
−
model 969-9333 for CFF 8” flange.
They typically reduce the vibration transmitted from
the Turbo-V300HT pump to the system by a factor
of 20.
Please refer to the relevant instruction manual.
Содержание Turbo-V300HT
Страница 2: ...Turbo V 300HT...