Reference
Page 58
User Manual Micro-GC
Varian, Inc.
GENERAL ERROR STATE
The General Error State as stored in UserDataStore address 1219 is composed of the following
items:
The error must be handled as ‘CLNNN’ in which C=error class, L=location and NNN=error
number or event number.
The Error class contains of the following values: 0=diagnostic error, 1=advisory error, 2=critical
error, 3=fatal error.
There are five locations: 0=main board, 1=channel 1, 2=channel 2, 3=channel 3 and 4=channel
4.
Error number NNN
Error Description
1
Init passed (event)
2 Pressure
restored
3
Start flush cycle
4 Flush
cycle
passed
5 TCD
calibrating
6
Too low pressure
7 Pressure
fault
8
Low battery 1
9
Low battery 2
10
Sample line sensor fault
11
Sample line temperature fault
12
Injector temperature fault
13
Column temperature fault
14
TCD Temperature limit activated
15
EDS logging error
16
Low power supply
17 Injector
sensor
fault
18
Column temperature sensor fault
19
TCD control error
20 TCD
calibration
failed
21 Hardware
reset"
22 Pressure
too
high
23 Initialization
error
24
Internal communication error
25
Instrument EDS incorrect
26 EDS
incorrect
27
Internal power failure
28 Flush
cycle
aborted
29
GC module changed
30
TCD Gain calibrated
31 TCD
Offset
calibrated
32 Null
String