Troubleshooting
4-2
Table 4-1 Troubleshooting
(Cont’d)
Fault
Probable Cause
Corrective Action
Poor ultimate
Contaminated pump fluid
Examine and clean pump;
pressure
replace fluid.
Low heat input
Check voltage. Check for
continuity, burned-out element,
poor thermal contact.
Inadequate cooling water flow
Check water pressure. Check
tubing for obstructions and
backpressure.
Excessive or too cold cooling water
Check temperature. Adjust flow.
High forepressure
Check for leak in foreline, poor
mechanical pump performance,
breakdown of mechanical
pump fluid.
Water in quick-cool coil
Check and remove cause.
Low heat input
Check heaters.
Low fluid level
Add fluid.
Malfunctioning pump assembly.
Check and repair or
Improperly located jets. Damaged
replace.
jet system.
Inlet pressure surges
Incorrect heater input
Check and correct.
Fluid outgassing
Condition fluid by operating the
pump for a few hours.
Leak in system ahead of pump inlet
Check and correct.
High chamber
Forepressure too high
Check for leak in foreline, poor
contamination
mechanical pump performance,
breakdown of pump fluid, and
incorrect valve operation.
Prolonged operation at high
Review procedures.
throughput at pressure above10
-3
Torr
Improper system operation and
Review procedures.
air release procedures
Pump will not start
Safety circuits and/or protective
Check utilities, flow devices
prevent contactor from
switches, interlocks. Check
staying closed
thermostat operation.
Содержание HS-20
Страница 18: ...Installation 1 3 Figure 1 2 Outline Drawing HS 20 Diffusion Pump with ASA Flanges ...
Страница 24: ...Installation 1 9 Figure 1 4 200 240 400 430 440 480 V Wiring Diagram HS 20 Diffusion Pump ...
Страница 25: ...Installation 1 10 Figure 1 5 380 415 V Wiring Diagram HS 20 Diffusion Pump ...
Страница 35: ...Maintenance 3 6 Figure 3 2 20 inch Jet Assembly ...
Страница 38: ...Maintenance 3 9 Figure 3 4 Heater Element Assembly NOTE ________ Tighten clamping nuts to 250 in lbs ...
Страница 41: ...Maintenance 3 12 ...