Model 990 dCLD II Component Leak Detector
2-6
2.2.3 Split Flow Dual Mechanical Pump Configuration
Figure 2-3 Split Flow Dual Mechanical Pump Configuration
Note: Foreline and Cal Leak valves are under the control of the 990 dCLD II.
Table 2-5 Split Flow Dual Mechanical Pump System Valve Sequence
State
Vent
Valve
Rough
Valve
Test
Valve
Remarks
Pre-Vented
Closed
Closed
Closed
The Pre-Vented state allows time to ensure that all valves
are closed before any open. The time spent in the
pre-vented state depends on valve actuation time.
Vented
Opened
Closed
Closed
The Vented state places the fixture at atmosphere.
Pre-Roughing Closed
Closed
Closed
The Pre-Roughing state allows time to ensure that all
valves are closed before any open. The time spent in the
pre-roughing state depends on valve actuation time.
Roughing
Closed
Opened
Closed
The Roughing state brings the fixture to vacuum.
Test
Closed
Opened
Opened
The Test Valve re-opens at the appropriate foreline
pressure, exposing the fixture to the foreline.
Sp
e
c
Tu
rb
o
Cal Le
ak
Cal
Leak
Foreline
Turbo Spec Tube Assembly
Length is Proportional
Test Valve
Rough Valve
Vent Valve
Fixture
(in box)
Backing
Primary
Pump
Roughing
Primary
Pump
Keep Distance as
Short as Possible
to Sensitivity
Tu
be
Содержание 990 dCLD II
Страница 4: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank D R A F T 4 2 0 0 7 ...
Страница 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 161: ......