www.vareximaging.com
37
1313DXT-I/1515DXT-I/2520DX-I/3030DX-I X-ray Detectors
U
SER
M
ANUAL
10.5.2
Offset Calibration Steps
1
Select the
desired mode
from the Mode Drop Down Menu.
Figure 37
Select the Desired Mode
2
Click the
Offset Calib button
or
click
Acquisition
from the menu bar and click on
Offset
Calibration
.
Figure 38
Offset Calibration Initiation
Note
A different offset reference image is necessary for each operating mode,
therefore it is important to update the offset data for each of the operating
modes.
Note
Perform an offset calibration in each mode when a gain calibration is
performed.
Caution
Wait at least 20 seconds after an X-ray exposure to initiate Offset
Calibration due to lag in the detector, this delay avoids a latent image in
the Offset Reference Image.