Guidance and manufacturer’s declaration – Electromagnetic immunity – Power and Signal PORTs

 

The AmpMC 3-field is intended for use in the Electromagnetic environment specified below. The customer or user of the 
AmpMC 3-field should assure that it is used in such an environment. 

Immunity test

 

Compliance Test level

 

Electromagnetic environment - guidance

 

Electrostatic 
discharge (ESD)  
IEC 61000-4-2 

± 8 kV contact 
± 2 kV, ± 4 kV, ± 8 kV, ± 15 
kV air 

Floors should be wood, concrete or ceramic tile. If floors are covered 
with synthetic material, the relative humidity should be at least 30 %. 

Electrical fast 
transient/burst  
IEC 61000-4-4 

± 2 kV AC and DC power 
ports 
± 1 kV signal ports 
100kHz repetition frequency 

Mains power quality should be that of a typical commercial or 
hospital environment. 

Surges 
Line-to-line 
IEC 61000-4-5 

Not Applicable 

The AmpMC 3-field is intended to be supplied by a secondary 
IEC60601-1 compliant AC or DC power supply. 

Surges 
Line-to-ground 
IEC 61000-4-5 

Not Applicable 

Conducted 
disturbances 
induced by RF fields 
IEC 61000-4-6 

 

3 V 

a)

  

0,15 MHz – 80 Mhz 

 
6 V 

a)

 in ISM bands between 

0,15 MHz and 80 MHz 

b) 

 
80 % AM at 1 kHz 

Field strengths from fixed RF transmitters, as determined by an 
Electromagnetic site survey 

c)

, should be less than the compliance 

level in each frequency range.

 

Interference may occur near equipment marked with the following 
symbol: 

 

 

Voltage dips 
IEC 61000-4-11 

Not Applicable 

The AmpMC 3-field is intended to be supplied by a secondary 
IEC60601-1 compliant AC or DC power supply. 

Voltage interruptions  
IEC 61000-4-11 

Not Applicable 

a)

  r.m.s. before modulation is applied 

b)

 The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 

MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 
MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 
10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 
29,7 MHz and 50,0 MHz to 54,0 MHz. 

c)

 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To 
assess the Electromagnetic environment due to fixed RF transmitters, an Electromagnetic site survey should be 
considered. If the measured field strength in the location in which the AmpMC 3-field is used exceeds the applicable RF 
compliance level above, the AmpMC 3-field should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as re-orienting or relocating the AmpMC 3-field. 

NOTE  

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 
structures, objects and people 

 

 

 

Содержание Claymount 1001

Страница 1: ...TM20514 10 Revision 3 0 www vareximaging com 0120 AmpMC 3 field Series Intended Part numbers as listed in Table 1 Technical Manual...

Страница 2: ...1 Check according to the build in LED s 10 3 2 2 Generator Switch off check 11 Necessary recurrent testing 11 4 Mains isolation 12 5 Service maintenance and cleaning 12 Safety precautions 12 Cleaning...

Страница 3: ...out Mechanical 13 Figure 4 DIL switch location 13 Figure 5 The connection against GND in active H mode 15 Figure 6 The connection against GND in active L mode 15 All rights reserved Although this guid...

Страница 4: ...rmance Varex Imaging Nederland B V hereby declares that this product is in conformity with the essential requirements and provisions as set forth in European Union Council Directive 93 42 EEC concerni...

Страница 5: ...pment IEC 60601 2 54 Medical electrical equipment Part 2 54 Particular requirements for the basic safety and essential performance of X ray equipment for radiography and radioscopy General warnings ca...

Страница 6: ...ON CABLE SSMC 15M EXTENSION CABLE SSMC 18M EXTENSION CABLE SSMC 20M EXTENSION CABLE SSMC 25M Extension cables for use between the AmpMC and HV generator EXTENSION CABLE PREAMP 10M SubD 9 pin male SubD...

Страница 7: ...l stop generating X rays automatically when the demanded radiation dose on the detector is reached Principle of operation The AmpMC main functions are The conversion of electrical current from a Solid...

Страница 8: ...apply see the documentation for the X ray system The AmpMC does not add new contraindications on top of these Overview of the device Table 1 AmpMC specifications Model Dimensions mm Support informatio...

Страница 9: ...2 VDC 0 1 A 12 VDC 0 1 A 15 VDC 0 1 A 15 VDC Output signal Ramp Positive or negative ramp 0 to 10 V 12 V 0 to 12 V 15 V Max Load 47 k Applicable standards IEC60601 1 2005 edition 3 0 C1 C2 IS1 IS2 A11...

Страница 10: ...e AmpMC always outside of the active image area The position of the AmpMC is not critical and can be mounted anywhere on the system If the cable lengths are inadequate see the accessories list in 1 7...

Страница 11: ...off check Without a patient or phantom a radiation exposure with 80kV 100mA 21 s is released The Automatic Exposure Control must terminate the exposure in less than 100 ms Necessary recurrent testing...

Страница 12: ...according to the described test procedure 5 4 Safety precautions When there is structural damage to the housing or cable of the device label the device as out of order and have the device repaired pr...

Страница 13: ...mpMC 3 field Series Revision 3 0 Date of release 2019 01 09 13 24 www vareximaging com 6 Device Data Figure 2 Image of an AmpMC 1001 1006 1008 3 field integrating amplifier Figure 3 Lay out Mechanical...

Страница 14: ...on of setting the DIL switches ON Switch closed OFF Switch open X Any position The position of the DIL switches on the PCB board is shown in Figure 4 The sensitivity step is selected by means of DIL s...

Страница 15: ...active H mode for reset exposure input is performed by means of switch 3 on DIL switch S2 S2 3 Interface type Function reset exposure Setting code OFF Active H H Exposure L Reset 0 ON Active L L Expo...

Страница 16: ...puts The plug pinout on the SUB D plug for the measuring field selection may be assigned by means of DIL switch S3 S3 5 S3 6 S3 7 S3 8 Field I Field II Field III Setting code ON OFF OFF x Pin 3 Pin 2...

Страница 17: ...performance of this equipment could result WARNING Not taking EMC measures into account on the wiring may result in increased EMISSIONS or decreased IMMUNITY IEC60601 1 2 2014 must be followed for bei...

Страница 18: ...lass B is normally required this equipment might not offer adequate protection to radio frequency communication services The user might need to take mitigation measures such as relocating or re orient...

Страница 19: ...460 FM 5 kHz deviation 2 0 3 28 1 kHz sine 710 704 787 LTE Band 13 17 Pulse Modulation b 0 2 0 3 9 745 780 217 Hz 810 800 960 GMS 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse Modulation b 2 0...

Страница 20: ...nce may occur near equipment marked with the following symbol Voltage dips IEC 61000 4 11 Not Applicable The AmpMC 3 field is intended to be supplied by a secondary IEC60601 1 compliant AC or DC power...

Страница 21: ...e Symbol Explanation Manufacturer Date of manufacture Catalogue number Serial number 0120 CE mark directive 93 42 EC conformity assessment by notified body 0120 Follow the instructions for use Reading...

Страница 22: ...Technical Manual TM20514 10 AmpMC 3 field Series Revision 3 0 Date of release 2019 01 09 22 24 www vareximaging com...

Страница 23: ...com Varex Imaging Philippines Inc Philippines t 63 49 5024 520 t 63 49 5024 521 f 63 49 5024 500 f 63 2 8076 472 Philippines CNC vareximaging com Varex Imaging Technology Beijing Co Ltd China t 86 10...

Страница 24: ...Technical Manual TM20514 10 AmpMC 3 field Series Revision 3 0 Date of release 2019 01 09 24 24 www vareximaging com...

Отзывы: