Technical Manual: TM20514-10
AmpMC 3-field Series
Revision: 3.0
Date of release: 2019-01-09
15/24
www.vareximaging.com
Ramp Polarity settings
The polarity of the output voltage ramp is selected by means of switches 1 and 2 of DIL-switch S2.
S2.1
S2.2
Ramp polarity
Setting
code
OFF
OFF
Output open
0
ON
OFF
Positive ramp
1
Standard version
OFF
ON
Negative ramp
2
ON
ON
Output Shorted, not allowed
-
Inputs reset / exposure settings
The selection of active L or active H mode for reset / exposure input is performed by means of switch
3 on DIL-switch S2.
S2.3
Interface
type
Function reset / exposure
Setting
code
OFF
Active H
H=Exposure
L=Reset
0
ON
Active L
L=Exposure
H=Reset
1
Standard version
Inputs for measuring field selection
The choice of active L or H mode for the field selection inputs is performed by means of switch 4 in
DIL-switch S2.
S2.4
Interface
type
Function Field selected
Setting
code
OFF
Active H
H=Field selected
L=Field off
0
ON
Active L
L=Field selected
H= Field off
1
Standard version
Figure 6 The connection against GND in active L mode
Figure 5 The connection against GND in active H mode