9. HYGIENE
Storage
Store Filling Chamber, Mouthpiece, Lip
Piece and Valve Balloon in a dry, dust-
free place away from contamination.
9.5. Disassembly and Assembly
Filling Chamber and Filling Cham-
ber Cap
Please allow the Filling Chamber and
the Cap to cool down before disas-
sembling them.
All component parts of the Filling
Chamber must be cleaned at the in-
tervals prescribed in Chapter 9.4.
“Reusability of the Filling Chamber”,
page 41. Disassemble the Filling
Chamber as follows:
Turn the Filling Chamber Cap coun-
terclockwise in order to remove it from
the Filling Chamber.
Disassembling the Filling Cham-
ber Cap
Use, e.g., the handle of the Cleaning
Brush to press the upper Screen down
in order to push it out. No Screens are
needed when using dronabinol.
Push the cap cylinder downwards out
of the cap housing. At the same time
the cap ring will be stripped off.
Push out the upper Screen with the help of
the Cleaning Brush
Push the cap cylinder out of the cap
housing
Remove Filling Chamber Cap
9. HYGIENE
Disassembling the Filling Chamber
Push the lower Screen or the Liquid
Pad upwards out of the Filling Cham-
ber Cylinder.
Push the Filling Chamber Cylinder
upwards out of the Filling Chamber
Housing.
Push the Filling Chamber Cylinder out of the
Filling Chamber Housing
When the Filling Chamber Cylinder is
pushed out, the three Filling Chamber
Clips are loosened and fall out of the
Filling Chamber Housing.
Drying
All component parts of the Filling
Chamber must be completely dry be-
fore reassembly.
Put all component parts of the Filling
Chamber on a clean, dry, and absor-
bent pad and allow them to dry thor-
oughly. A hair dryer may be used for
faster drying. Do not dry in wet rooms
(e.g. not in the bathroom).
Assembly of Filling Chamber Cap
Slide the cap cylinder into the cap
housing until it clicks, then pull the cap
ring from above over the cap cylinder
to the cap housing.
When using hemp flowers, push the
upper Screen from below into the cap
cylinder until it clicks into the desig-
nated notch.
When using dronabinol solution, the
two Screens are not needed. Only the
Liquid Pad is inserted instead of the
lower Screen.
Lower Screen
Filling Chamber
Cylinder
Filling Chamber
Housing
3 pcs. Filling
Chamber Clip