V&P Scientific VP 548-US Скачать руководство пользователя страница 1

V&P Scientific, Inc.  

Technical Note 317 

 

 
 
 

Technical Note 317 

 

Operation Manual for 

VP 548-US

 

Ultrasonic Wash

 

Station 

 

Introduction 
 

This is the user manual for the 

VP 548-US

 

ultrasonic wash station.  This wash station is designed for 

cleaning plastic tips, needles and pins used in liquid handling applications.  It has a standard SBS 
microplate footprint and is designed to fit on the deck of most liquid handling robots.  It uses high 
frequency ultrasound to generate an oscillating cavitation field in a liquid bath. This is an effective 
means of removing organic material that is attached to hard surfaces.  The 

VP 548-US 

is equipped 

with a weir system to provide a constant level of wash liquid and to prevent overflows. 

 
Safety 

 

1. Do not attempt to open the ultrasonic tank (which contains the ultrasonic generator) or control and 
pump unit. There are high voltages present and there are no user serviceable parts.  

 

2. Water is an effective cleaning agent in an ultrasonic bath. You may add to this a surfactant or a 
miscible organic solvent. The solution must be at least 50% water.  
 

Note:

  The tank has not been tested with a range of organic solvents. Operators MUST ensure that 

the use of solvents in the wash station has been sufficiently investigated such that a hazard is not 
caused. Operators must also ensure that the solvent mixture does not damage tubing, electrical 
cables or fittings. V&P recommends that water should be used wherever possible. V&P accepts no 
responsibility for damage or loss if solvents other than water are used.  
 
 

 

Содержание VP 548-US

Страница 1: ...nstant level of wash liquid and to prevent overflows Safety 1 Do not attempt to open the ultrasonic tank which contains the ultrasonic generator or control and pump unit There are high voltages presen...

Страница 2: ...erate in a hazardous environment or where flammable vapor exists The ultrasonic wash station is not rated or certified to work in hazardous environments 7 Clean carefully only with water and detergent...

Страница 3: ...r Converter from 115VAC to 220 240VAC E USB Computer Connection Cable F Tubing to fill and drain Ultrasonic Tank G USB memory stick with software Figure 2 Rear panel connections Figure 3 Front panel c...

Страница 4: ...t to a compatible computer see Figure 5 The Control and Pump Unit should be connected to an AC power outlet 220 240 Vac 50 60 Hz with a power rating of at least 150 VA Mount the ultrasonic tank on the...

Страница 5: ...ed soil printed on both sides The strip is placed within a holder and simulates soiling on surgical instruments described in BS2745 and prEN15883 We recommend the use of Browne STF Load Check ref 2315...

Страница 6: ...each element of the VP 548 US the ultrasonic power RF the fill and empty pumps The fill pump can be reversed to empty the tank Caution this could lead to contaminating the clean wash liquid reservoir...

Страница 7: ...s to operate 1 Check all electrical connections 2 Check status of blue Power Indicator Light on the Control and Pump Unit If this is off check power connections 4 All repairs are on a return to manufa...

Страница 8: ...function switches Pump 1 fill and Pump 2 Drain on for 10 seconds switches them off and then the ultrasound on for 20 seconds Figure 9 Two Command Line Arguments Low level command set Function Instruct...

Отзывы: