11.2 Connection settings
Current outputs:
Select
Settings
>
Connection settings
>
Current outputs
.
Current output 1/2:
Configure the consistency current output to Output signal 1.
Configure temperature, pressure or measurement attenuation to Output signal 2.
Min, Max:
Enter values for the low (4 mA) and high (20 mA) consistency values. Info
button shows the default range.
Damping sec:
Set output signal damping, 1 to 60 seconds.
Alarm type:
Analog output signal level indicating a fault. Alternatives: Low, High,
Freeze (= analog output is frozen to the level it had when the error occurred), Preset (=
user specified current value) or NONE (= error situation is disregarded, result
calculation continues).
HART address:
Default value is 0 (zero), i.e. normal installation with a single
transmitter, analog output signal (4–20 mA) is active. Values 1–63 define a fixed current
output and all communication with the device takes place digitally.
Configuration | 37 (93)
K17122 V1.1 EN - 1 - 06/2020