Rear Panel
2-6
RapidVac™ Smoke Evacuator Service Manual
The Generator Interlink Cable is compatible with the Valleylab Force FX™,
ForceEZ™, and ForceTriad™ generators. It attaches to a generator’s RF
activation port and to the smoke evacuator’s Remote Activator jack. When the
generator is activated, the smoke evacuator automatically initiates the pre-set
suction.
The Remote Switch Activator is a programmable switch that plugs into the smoke
evacuator jack and is attached to an electrosurgical generator. The remote switch
is programmed to recognize the audible tones used by the generator. When the
tone for cut or coag sounds, the remote switch automatically initiates the pre-set
suction.
EPROM Jack (Covered)
The EPROM jack located under the secured cover is to be used only by qualified
Valleylab service technicians to update the smoke evacuator’s software.
Equipotential Grounding Lug
The equipotential grounding lug provides an alternate rout for electrical energy
back to ground when attached to an earth-ground cable.