DID-UM-002
Val Controls A/S
•
Limfjordsvej 3
•
DK-6715 Esbjerg N
•
Tel. +45 7547 0600
•
www.valcontrols.com
Revision: 10
7/74
4.3
Buttons - menus
Four menus can be accessed from the main screen by pressing each of the buttons or using
MTControl.
The F model only has one button. Pressing it for 3 seconds will initiate the auto calibration. The
tests and auto calibration can be aborted by pressing the button for one second.
4.3.1
Test menu
Test Menu
Description
1
Start PST
Start a partial stroke test
2
Start FST
Start a full stroke test
3
Start SOT
Start a solenoid test
4.3.2
Diagnostic menu
Diagnostic Menu
Description
1
PST
View the reference and latest partial stroke test values and limits
Date
Date the test was performed
Time
Time the test was performed
Last test result
Result of the latest test
Start pressure
Start pressure
React
SOV reaction time
Breakaway time
Breakaway time
Breakaway pressure Breakaway pressure
PST time
PST time
Press
PST pressure
PrsTi
PST pressurising time
RetTi
PST return time
FilTi
Actuator fill time
Total time
PST total time
Max. travel
PST maximum travel
Min. pressure
PST minimum pressure
Enable errors
Enable errors
Enable warnings
Enable warnings
2
FST
View the reference and latest full stroke test values and limits
Date
Date the test was performed
Time
Time the test was performed
Last test result
Result of the latest test
Start pressure
Start pressure
React
SOV reaction time
Breakaway time
Breakaway time
Breakaway pressure Breakaway pressure
Breakaway pos.
limit
Breakaway position error limit
FST time
FST time
DePrs
FST depressurising time
PrsTi
FST pressurising time
RetTi
FST return time
FilTi
Actuator fill time
Total time
FST total time
Enable errors
Enable errors
Enable warnings
Enable warnings
3
SOT
View the reference and latest solenoid valve test values and limits
Date
Date the test was performed
Time
Time the test was performed