Chapter 11 ____________________________________________________ Technical Specifications
VAISALA______________________________________________________________________ 177
- When the analog output option is applied, digital communication is not available.
General Conditions
- Due to the measurement frequency used in the sonic transducers, RF interference in the
200 ... 400 kHz range can disturb wind measurement.
- Any temporary element or object, such as snow, ice, or a bird, blocking the measurement path
between the ultrasonic transducer heads can affect the wind measurement accuracy or even
invalidate the output data.
- Extreme operating conditions can alter sensor performance temporarily.
Table 36
Analog mA Output Options
Parameter
Description/Value
Wind speed
0 ... 20 mA or 4 ... 20 mA
Wind direction
0 ... 20 mA or 4 ... 20 mA
Load impedance
200
Ω
max
Update interval
4 Hz max
Table 37
General Conditions
Property
Description/Value
Housing protection class
IP65 (without mounting kit)
IP66 (with mounting kit)
Temperature
operative
storage
-52 ... +60 °C (-60 ... +140 °F)
-60 ... +70 °C (-76 ... +158 °F)
Relative humidity
0 ... 100 %RH
Pressure
600 ... 1100 hPa
Wind
0 ... 60 m/s
CAUTION
Make sure that you power the sensor after installation. Storing the
sensor outdoors without a proper package or not powering up after
installation can affect the sensor’s expected lifespan.
NOTE
Select a heated sensor model if you operate the sensor in humid
conditions.
Содержание WXT536
Страница 1: ...USER S GUIDE Vaisala Weather Transmitter WXT530 Series M211840EN C...
Страница 235: ...www vaisala com M211840EN...