Chapter 8 ____________________________________________ Sensor and Data Message Settings
VAISALA______________________________________________________________________ 145
Setting Fields
[R]
=
Parameter selection: This field consists of 16 bits
defining the supervisor parameters included in the
data messages. The bit value 0 disables and the bit
value 1 enables the parameter.
Bits 1-8 determine the
parameters included in the
message obtained with the
following commands:
-
ASCII:
aR5 and ar5
-
NMEA 0183:
$--WIQ,XDR*hh
-
SDI-12:
aM5, aMC5, aC5, and
aCC5
-
SDI-12 continuous:
aR5 and
aRC5
1st bit (most left)
Th Heating temperature
2nd bit
Vh Heating voltage
3rd bit
Vs Supply voltage
4th bit
Vr 3.5 V reference voltage
5th bit
Id Information field
6th bit
spare
7th bit
spare
8th bit
spare
&
delimiter
Bits 9-16 determine the
supervisor parameters
included in the composite data
message obtained with the
following commands:
-
ASCII:
aR0 and ar0
-
NMEA 0183:
aR0, ar0
-
SDI-12:
aM, aMC, aC, and aCC
-
SDI-12 continuous:
aR and
aRC
9th bit
Th Heating temperature
10th bit
Vh Heating voltage
11th bit
Vs Supply voltage
12th bit
Vr 3.5 V reference voltage
13th bit
Id Information field
14th bit
spare
15th bit
spare
16th bit (most right)
spare
[I]
=
Update interval: 1 ... 3600 seconds. When the
heating is enabled the update interval is forced to 15
seconds.
[S]
=
Error messaging: Y = enabled, N = disabled
[H]
=
Heating control enable: Y = enabled, N = disabled
Heating enabled: The control between full and half
heating power is on as described in
.
Heating disabled: Heating is off in all conditions.
<cr><lf>
=
Response terminator