5. Troubleshooting
5.1 Troubleshooting WAT12
Table 9 Common problems and their remedies
Problem
Probable cause
Remedy
Data is not received by the data
collecting system.
Improper or loose connections. Check wiring and tighten the
screw terminals.
Shaft heating of the sensors is
not working.
Improper or loose connections. Check wiring and tighten the
screw terminals.
Heating is not connected.
Check the connections.
The thermostat switch is failing. Replace the thermostat switch.
Error messages
If the wind speed or direction sensor is removed (output current is 0), the values in the
messages are replaced by slashes and the I/O error bit of status is set to 1.