6. Technical Data
6.1 PTB210 specifications
Table 5 PTB210 measurement performance
Property
Description/Value
Pressure range
Serial output
500 … 1100 hPa
50 … 1100 hPa
Analog output
500 … 1100 hPa
600 … 1060 hPa
800 … 1060 hPa
900 … 1100 hPa
Serial output, accuracy (hPa)
Pressure range
500 … 1100
50 … 1100
Class A
Class B
Non-linearity
± 0.10
± 0.15
± 0.20
Hysteresis
± 0.05
± 0.05
± 0.10
Repeatability
± 0.05
± 0.05
± 0.10
Calibration uncertainty
± 0.07
± 0.15
± 0.20
Accuracy at +20 °C
(+68 °F)
± 0.15
± 0.20
± 0.35
Temperature
dependency
± 0.20
± 0.20
± 0.40
Total accuracy
-40 … +60 °C
(-40 … +140 °F)
± 0.25
± 0.30
± 0.50
Long-term stability
(hPa/year)
± 0.10
± 0.10
± 0.20
Analog output, accuracy
Non-linearity
± 0.20 hPa
Hysteresis
± 0.05 hPa
Repeatability
± 0.05 hPa