Table 2 PTB210 wiring with RS-232C

Wire color

Signal

Grey

RX

Green

TX

Blue

Ground

Pink

Supply voltage (5 ... 28 V DC)

Brown

Not connected

White

Not connected

Yellow

Power down mode (TTL level: 0 V = off, 5 V = on)

Table 3 PTB210 wiring with RS-485/RS-232C interface

Wire color

Signal

Grey

RX (with RS-232C)

Green

TX (with RS-232C)

Blue

Ground

Pink

Supply voltage (8 ... 18 V DC)

Brown

RS-485 -

White

RS-485 +

Yellow

Power down mode (TTL level: 0 V = off, 5 V = on)

The 2 output protocols cannot be used simultaneously. Connect either RX/TX or
RS-485 wires.

Table 4 PTB210 with RS-232C/TTL interface

Wire color

Signal

Grey

RX (with RS-232C)

Green

TX (with RS-232C)

Blue

Ground

Pink

Supply voltage (5 ... 28 V DC)

Brown

RXD (TTL level)

White

TXD (TTL level)

Yellow

Power down mode (TTL level: 0 V = off, 5 V = on)

PTB210 Series User Guide

M210669EN-E

12

Содержание PTB210 SERIES

Страница 1: ...M210669EN E User Guide Digital Barometers with Serial Output PTB210 Series...

Страница 2: ...al English versions In ambiguous cases the English versions are applicable not the translations The contents of this document are subject to change without prior notice Local rules and regulations may...

Страница 3: ...1 6 AVRG defining averaging 16 4 2 Operating commands 16 4 2 1 P outputting single message 16 4 2 2 BP continuous output 17 4 2 3 UNIT selecting pressure unit 17 4 2 4 FORM output with or without uni...

Страница 4: ...Recycling 31 PTB210 Series User Guide M210669EN E 2...

Страница 5: ...List of figures Figure 1 PTB210 with standard accessories 7 Figure 2 BAROCAP pressure sensor 7 Figure 3 PTB210 mounting options 9 Figure 4 PTB210 dimensions 28 List of figures 3...

Страница 6: ...with RS 485 RS 232C interface 12 Table 4 PTB210 with RS 232C TTL interface 12 Table 5 PTB210 measurement performance 25 Table 6 PTB210 operating environment 26 Table 7 PTB210 inputs and outputs 26 Ta...

Страница 7: ...erious hazard If you do not read and follow instructions carefully at this point there is a risk of injury or even death WARNING Caution warns you of a potential hazard If you do not read and follow i...

Страница 8: ...l size facilitates easy installation and a long lifetime for the barometer PTB210 is ideal for use in applications such as weather stations data buoys ships airports and agrology It is also an excelle...

Страница 9: ...veloped by Vaisala for barometric pressure applications The Vaisala BAROCAP sensor provides excellent hysteresis and repeatability characteristics and outstanding temperature and long term stability P...

Страница 10: ...carefully matched together to minimize the temperature dependence and to maximize the long term stability The BAROCAP pressure sensor is designed to achieve zero temperature dependence at 1000 hPa and...

Страница 11: ...aded screws the O ring sealing the attachment A Fix the supplied attachment plate to a desired place and attach the barometer on the plate with the supplied screws supplied B Attach PTB210 to a standa...

Страница 12: ...itting You can also use a static pressure head SPH10 20 to protect the pressure channel from rain and minimize the effect of wind 3 1 1 Static pressure head The pressure fitting supplied with the tran...

Страница 13: ...ferent serial output interfaces available for PTB210 digital barometers RS 232C RS 485 232C RS 232C TTL The connections depend on the output protocol in use Vaisala recommends isolating the free wire...

Страница 14: ...TX with RS 232C Blue Ground Pink Supply voltage 8 18 V DC Brown RS 485 White RS 485 Yellow Power down mode TTL level 0 V off 5 V on The 2 output protocols cannot be used simultaneously Connect either...

Страница 15: ...The 2 output protocols cannot be used simultaneously Connect either RX TX or RXD TXD wires More information RON ROFF setting RS 485 terminating resistor on off page 18 Chapter 3 Installation 13...

Страница 16: ...he code directs the command to that specific instrument This prevents any inconsistency between the barometers and the host system A barometer reacts only to a command with its ID or the one without t...

Страница 17: ...ode is 0 When several barometers are on the same line use the ID code to direct the command to a specific instrument separately The barometer handles the ID code as characters for example 0001 01 and...

Страница 18: ...RG is 0 The command sets the number of measurement points that are used for the calculating an average pressure reading This is not a moving average Example If the number of measurements per minute is...

Страница 19: ...r command invisible 997 99 998 01 998 01 998 02 cr command invisible 4 2 3 UNIT selecting pressure unit UNIT x x 0 hPa 1 mbar 2 inHg 3 psia 4 torr 10 bar 5 mmHg 6 kPa 7 Pa 8 mm H2O 9 in H2O 10 bar The...

Страница 20: ...activates the multi point corrections MPCOFF The command deactivates the multi point corrections 4 2 7 CALD setting calibration date CALD stores the last calibration date in the memory of the baromete...

Страница 21: ...t the end of the serial bus there must be PTB210 a dynamic line adapter or a line master RON The RON command sets the terminating resistor on ROFF The ROFF command sets the terminating resistor off 4...

Страница 22: ...ATE 1999 06 10 ID CODE 10 SERIAL NUMBER T00100004 MULTIPOINTCORR ON MEAS PER MINUTE 60 AVERAGING 0 PRESSURE UNIT hPa Pressure Min Max 500 1100 50 1100 for model 50 1100 LOW CURRENT MODE RS485 RESISTOR...

Страница 23: ...adjustments By comparing the output to a reference you can define the corrections required for readjustment Write down the previous multi point corrections To make the corrections use the Wizard soft...

Страница 24: ...ires See Table 2 page 12 1 Go to http www vaisala com PTB210Wizard PTB210_calibration_wizard zip downloads 2 Extract the zip file and doubleclick the exe file 3 Select Run Next Next Next Install Finis...

Страница 25: ...ons according to the following equation You can also change the settings on the left corr Preference Pmeasured The values are in hPa The reference value is as whole number and the correction with 2 de...

Страница 26: ...here is no password You can protect the new corrections with a password To enter a password select the Change 9 Type a new password and type it again when prompted Make sure you set the multi points c...

Страница 27: ...range 500 1100 50 1100 Class A Class B Non linearity 1 0 10 0 15 0 20 Hysteresis 1 0 05 0 05 0 10 Repeatability 1 0 05 0 05 0 10 Calibration uncertainty 2 0 07 0 15 0 20 Accuracy at 20 C 68 F 3 0 15 0...

Страница 28: ...f end point non linearity hysteresis error repeatability error and calibration uncertainty at room temperature 4 Defined as 2 standard deviation limits of temperature dependence over the operating tem...

Страница 29: ...rrent consumption shutdown mode 0 2 mA All models Max pressure 5 000 hPa absolute Pressure connector M5 10 32 internal thread Pressure fitting Barbed fitting for 1 8 in I D tubing Supply voltage rever...

Страница 30: ...4 30 EU RoHS Directive 2011 65 EU EMC compatibility EN IEC 61326 1 Electrical equipment for measurement control and laboratory use EMC requirement Basic environment CISPR 32 EN 55032 Class B Complianc...

Страница 31: ...turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase the separation...

Страница 32: ......

Страница 33: ...of the warranty for each product Technical support Contact Vaisala technical support at helpdesk vaisala com Provide at least the following supporting information as applicable Product name model and...

Страница 34: ......

Страница 35: ......

Страница 36: ...www vaisala com...