Register
Number
(Decimal)
Address
(Hexa
-
decimal
)
Register Description
Data Format
Bitmask
521
0208
hex
CH
4
measurement
status
16-bit signed
integer
0000
hex
: Unsupported
0001
hex
: Reading is not
reliable
0002
hex
: Under range
0003
hex
: Over range
0004
hex
: Noisy
0005
hex
: Value locked
0006
hex
: Calibration expired
0007
hex
: Sensor failure
0008
hex
: Measurement not
ready
522
0209
hex
CH
4
measurement
status (dry basis)
16-bit signed
integer
523
020A
hex
CO
2
measurement
status
16-bit signed
integer
524
020B
hex
CO
2
measurement
status (dry basis)
16-bit signed
integer
525
020C
hex
H
2
O measurement
status
16-bit signed
integer
526
020D
hex
T
df
measurement status 16-bit signed
integer
527
020E
hex
T
s
measurement status
16-bit signed
integer
528
020F
hex
Device status
16-bit signed
integer
0000
hex
: Critical failure
0001
hex
: Error
0002
hex
: Warning
0003
hex
: Notification
529
0210
hex
Clear error log: the error
log is cleared if 1 is
written, other values are
ignored. Register always
reads as 0.
Function
Table 16 Error Codes in Register 0200
hex
(Status Code Low)
Bitmask
Error Message
Severity
0000
hex
Firmware checksum mismatch.
Critical
0001
hex
Device settings corrupted.
Critical
0009
hex
Infrared source temperature too high.
Error
0010
hex
Infrared source failure.
Error
0011
hex
Infrared source failure.
Error
0012
hex
Supply voltage out of range.
Error
0013
hex
Internal voltage out of range.
Error
Appendix A
81
Содержание MGP261
Страница 8: ...MGP261 User Guide M212273EN A 6...
Страница 50: ...4 Store the setting with Save MGP261 User Guide M212273EN A 48...
Страница 86: ...MGP261 User Guide M212273EN A 84...
Страница 88: ...MGP261 User Guide M212273EN A 86...
Страница 89: ......
Страница 90: ...www vaisala com...