3.5 Finalizing installation
PULL INSTRUMENT OUT
BEFORE CLOSING VALVE
≥
4 mm2
2
1
1
Safety pin: in ball valve installations, lock the handle of the ball valve in the open position
with the safety pin to prevent damage caused by closing the valve with the instrument
inside.
2
Probe grounding terminal.
1. Close the connection box and tighten the captive Allen screws on the connection box
flange. Tighten the cable glands, and then tighten the probe to final tightness on the
installation port by turning the tightening nut with a 50 mm (1.97 in) wrench.
2. Connect the MGP260 grounding terminal to the grounding rail of the installation site with
a
≥
4 mm
2
wire.
3.
Ball valve installations only:
lock the handle of the ball valve in the open position with
the safety pin.
4. When done, switch on the power supply input.
3.6 Attaching flow-through adapter to MGP260
The MGP260 flow-through adapter 258877 is delivered assembled and is intended to be used
with the probe installed in an upright position. For an overview of the flow-through adapter,
Installation option: flow-through adapter installation (page 22)
To attach the adapter to MGP260:
Chapter 3 – Installation
31
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备 气象科学仪器设备
Tel:029-88186182 Web:www.xavery.cn Email:[email protected]
Содержание MGP260 Series
Страница 8: ...MGP260 User Guide M212504EN A 6 Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Страница 88: ...Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Страница 90: ...Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Страница 91: ...Tel 029 88186182 Web www xavery cn Email sxvery 163 com...
Страница 92: ...www vaisala com Tel 029 88186182 Web www xavery cn Email sxvery 163 com...