3. In
Analog input off/on
, set analog input on.
4. To power your analog input device through Indigo520, in
Analog input powering
, set
analog input powering on.
To see the changes you made, select
. Then use the left and right arrows at the
bottom of the touchscreen to navigate to the
I/O status
view.
6.5 Configuring relays on touchscreen
Relays are not available in transmitters that are powered with Power over
Ethernet (PoE).
The transmitter has 2 configurable relays. Both relays have configuration options for selecting
the parameter that is used to control the relay, activation triggers, and hysteresis.
Before you start, make sure that the probes you want to use are connected to the transmitter.
Figure 34
Relays
configuration menu on touchscreen,
Relay 2
tab active
1
The
Relay 1 … 2
tabs contain individual settings for each relay.
2
A circle next to the relay name indicates that the relay is on.
3
The
Test mode
tab contains the settings for testing each relay.
1. Select the configuration menu symbol at the upper right corner of the touchscreen.