![Vaisala Indigo 500 Series Скачать руководство пользователя страница 27](http://html2.mh-extra.com/html/vaisala/indigo-500-series/indigo-500-series_user-manual_3387257027.webp)
Table 4 AC power supply input terminals
Terminal
Function
Notes
Power supply ground
L
Line
100 … 240 VAC 50/60 Hz
N
Neutral
3.3.2.1 Preparing AC (mains) power cable
• Cable stripping tool
Prepare the AC (mains) power cable for connection to the power supply input terminal:
mm
[in]
55 [2.17]
45 [1.77]
8 [0.31]
1
2
3
Figure 9 Example of stripped AC (mains)
power cable
1
Line wire
2
Grounding wire
3
Neutral wire
1. Strip 55 mm (2.17 in) of the cable to expose the wires.
2. Cut off 10 mm (0.39 in) of the line and neutral wires. Leave the grounding wire 55 mm
(2.17 in) long.
Make sure that the grounding wire is longer than the line and
neutral wires. Under mechanical stress, the grounding wire must be the last
to disconnect from the protective ground terminal.
CAUTION!
3. Strip the ends of the individual wires to expose the conductors for a length of 8 mm
(0.32 in).
3.3.3 Relay output terminals and lead-through
Before connecting wires or cables, make sure that the transmitter is powered off.