7.1.2 Restoring Factory Adjustment
Always restore factory adjustment before entering a new adjustment. This prevents any
possible earlier adjustments having an effect on the new adjustment you make.
To restore factory adjustment:
1. Connect to the wireless configuration interface and open the
Calibration
menu.
2. Start the calibration mode with the
Start calibration
button.
3. On the
Calibration
tab, scroll down to the parameter you want to adjust (for example,
CO2 adjustment
) and select
Restore factory adjustment
.
4. Restore the factory adjustment with the
Restore factory adjustment
button for each
parameter separately as needed.
5. To verify that the factory adjustment was restored, check the adjustment data
information at the bottom of the
Calibration
tab view.
6. Close the calibration mode with the
Stop calibration
button.
7.1.3 Calibration PIN Code
Probe calibration can be locked and unlocked with a calibration PIN code in the
Settings >
Probe
menu. By default, the calibration PIN code is in place and calibration is enabled. Do
not remove the PIN code from the probe settings unless you need to block access to
calibration settings.
If the PIN code has been removed and you need to re-enter it, check the probe's
documentation for information on the code used in the probe model.
Chapter 7 – Calibration and Adjustment
47
Содержание Indigo 202
Страница 1: ...M211966EN A User Guide Indigo 202 Digital Transmitter Indigo 202...
Страница 22: ...Indigo 202 User Guide M211966EN A 20...
Страница 62: ...Indigo 202 User Guide M211966EN A 60...
Страница 70: ...Indigo 202 User Guide M211966EN A 68...
Страница 74: ...Indigo 202 User Guide M211966EN A 72...
Страница 75: ......
Страница 76: ...www vaisala com...