Changing the filter and sensor
1.
Unscrew the filter from the probe head.
2.
Remove the damaged sensor and insert a new one. Handle the
sensor by the plastic socket. DO NOT TOUCH THE SENSOR
PLATE.
3.
After sensor change the humidity calibration must be made
according to the instructions, see command
FCRH,
page 29 .
4.
Screw a new filter on the probe head. Take care to tighten the filter
properly, recommended force 130 Ncm.
New sensors and filters can be ordered from Vaisala, see list of
accessories on page 51.
Clean the sensor before storing the MMT318
and MMT317
Clean the used sensor with instrument air to blow out existing oil or
gently first flush with heptane (C7H14 16) and dry with instrument air
to prevent oxidation of the sensor. The oxidation of the sensor can
cause extended response times or drifting.
Factory calibration and repair service
The MMT318 and MMT317 are fully calibrated as shipped from
factory. The recommended calibration interval is 1 year. However,
calibration shall be done always when there is a reason to believe that