USER'S GUIDE____________________________________________________________________
38 __________________________________________________________________ M210474EN-B
Temperature calibration and
adjustment (in one point)
1.
Remove the transmitter unit from the mounting plate (see page
12) and press the adjustment button, see page 26 .
2.
Remove the probe filter and insert the probe head into the
reference temperature.
3.
Let the sensor stabilize.
4.
Type command
CT
and press
ENTER
.
CT
↵
5.
Type
C
and press
ENTER
a few times to check if the reading is
stabilized.
6.
When the reading is stabilized, give the reference temperature
after the question mark and press
ENTER
three times
.
>
ct
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? c
T : 16.06 Ref1 ? 16.0
Press any key when ready ...
T : 16.06 Ref2 ?
OK
>
OK
indicates that the calibration has succeeded
7.
Take the probe out of the reference conditions and replace the
filter. Take care to tighten the filter properly, recommended
force 130 Ncm.
8.
If needed, give the calibration information (date and text) to the
transmitter's memory, see the serial commands on page 29.
9.
Reset the transmitter by giving a command
RESET
. Transmitter
returns to normal mode.