Chapter 7 ___________________________________________________Calibration and Adjustment
VAISALA ______________________________________________________________________ 161
Temperature Adjustment
Using Display/Keypad
1. Press
the
ADJ
button on the motherboard to open the
ADJUSTMENT MENU
. If using a warmed probe for measuring,
probe heating will be interrupted when
ADJ
key is pressed. Wait
some time for the probe to reach ambient temperature.
2. Select
►
Adjust T measurement
, press
►
key.
3. Select
1-point/ 2-point adjustment
, press
START
.
4.
Remove the filter from the probe and insert the probe into the
reference temperature.
5.
Wait at least 30 minutes for the sensor to stabilize. Follow the
stabilization from the
GRAPH
display.
6. Press
READY
when stabilized. Give the reference temperature
using the arrow buttons.
When carrying out the 2-point adjustmentproceed to the next
adjustment point and carry out the procedure as described in the
previous item. Please, note that the difference between the two
temperature references must be at least 30 ºC.
7. Press
OK.
Press
YES
to confirm the adjustment.
8. Press
OK
to return to the adjustment menu.
9. Press
EXIT
to close the adjustment mode and return to the basic
display.
Using Serial Line
1. Press
the
ADJ
key on the motherboard to open the adjustment
mode. If using a warmed probe for measuring, probe heating will
be interrupted when
ADJ
key is pressed. Wait some time for the
probe to reach ambient temperature.
2.
Remove the probe filter and insert the probe into the reference
temperature.
3.
Enter the command
CT
or
(
CTA
for additional T probe)
and press
ENTER
.
>
ct
or for additional T probe:
>
cta