![Vaisala HUMICAP HMT310 Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/vaisala/humicap-hmt310/humicap-hmt310_user-manual_3340947046.webp)
USER'S GUIDE____________________________________________________________________
44 _______________________________________________________________ M210619EN-C
Other Commands
? Check Transmitter Settings
Syntax: ?<cr>
Syntax: ??<cr>
Use command ? to check the current transmitter configuration.
Command ?? is similar but can also be used if the transmitter is in
POLL state.
Example (factory default settings):
>?<cr>
HMT310 / 1.07
PRB serial nr : A0000000
Calibration : 2004-05-07
Cal. info : Vaisala/HEL
Output units : metric
Pressure : 1013.25 hPa
RS232 settings
Address : 0
Output interval: 0 S
Baud P D S : 4800 E 7 1
Serial mode : STOP
Analog outputs
Ch1 output mode: 0 ...20mA
Ch2 output mode: 0 ...20mA
Ch1 error out : 0.000mA
Ch2 error out : 0.000mA
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo : -40.00 'C
Ch2 T hi : 60.00 'C
>
ECHO Serial Bus Echo
Syntax: ECHO x<cr>
where
x
= ON/OFF (default = ON)
The command enables or disables echo of characters received over the
serial line.