HMI41 AND PROBES
Operating Manual
M210316EN-A
40
10.2
Checking the settings
Turn the HMI41 on and enter the setup mode (see Chapter 8). Make sure that
the following settings are correct:
setting
correct value
probe
AUT
(all probes)
or
1
(with HMP41/45) or
2
(with HMP46)
start (*)
1
baud
4.8
seri
E.7.1
calib
def
(**)
(*) values 2, 3 and 4 are for calibration cables
(**) when the
calib
setting is entered, the value is always “no”; factory
settings for calibration correction data can be returned by selecting “def”
If the settings are not correct, change them: in the setup mode, select the
setting with ENTER, change it with buttons
▲
or
▼
and acknowledge with
ENTER. If the error message still appears, write it down and contact Vaisala
or a Vaisala representative for further instructions.