6.4.3 CT command
Use the
CT
command for one-point or two-point serial line temperature calibration and
adjustment of the T sensor. The
CT
command changes the user adjustment parameters of
temperature sensor T.
You can use the
CT
command only in the adjustment mode.
Example of two-point adjustment:
>ct
T : 18.6038 1. ref ?
T : 18.6068 1. ref ?
T : 18.6098 1. ref ? 19.0
Press any key when ready ...
T : 49.5176 2. ref ?
T : 49.5176 2. ref ?
T : 49.5176 2. ref ? 50.0
OK
>
OK
indicates that the adjustment has been successful.
6.4.4 CTA command
Use the
CTA
command for two-point serial line temperature calibration and adjustment of the
additional temperature probe (Ta).
CTA
changes the user adjustment parameters of the
additional temperature probe. See also
and
.
CTA
command can only be used in adjustment mode.
Example of 2-point adjustment:
>cta
TA : 10.21 Ref1 ?
TA : 10.21 Ref1 ?
TA : 10.19 Ref1 ? 10.5
Press any key when ready ...
TA : 50.19 Ref2 ?
TA : 50.20 Ref2 ?
TA : 50.19 Ref2 ?
TA : 50.19 Ref2 ? 50.3
OK
>
OK
indicates that the adjustment has been successful.
Chapter 6 – Calibration and adjustment
75