4.10.3 FILT command
Use the
FILT
command to set the filtering level. The default setting is 0.8. The output is
calculated based on the following formula: [( new result * filt) + ( old result * (1.0 - filt))]. With
FILT
value 1, the probe only takes the latest measurement into account. With
FILT
value 0.1,
a new output is a combination of the previous output (90 %) and the latest measurement
(10 %).
FILT
xxx
Parameter
Description
xxx
0.1 … 1
1.0 = no filter
0.1 = about 16 moving average
Example
>filt
Filter : 0.800 ?
>
4.10.4 HELP command
Use the
HELP
command to list the available commands.