>xheat on
Extra heat : ON
>xheat off
Extra heat : OFF
>
To configure extra sensor heating, use the
XHEAT
command without parameters. Type the
values after the question mark. The following table lists the available ranges.
Available ranges
Description
0 ... 100 %RH (default:
95 %RH)
Extra heating RH limit (heating function starts up above the
setpoint)
0 ... 200 °C (default: 100 ºC)
Extra heating temperature
0 ... 255 s (default: 30 s)
Extra heating time
To see the current values, use the
XHEAT *
command.
After the defined
XHEAT
duration, there is an additional cooling time of 10
seconds before outputs are updated.
Example
>xheat *
Extra heat : OFF
RH limit : 95 %RH
Temperature : 100 'C
Duration : 30 s
>
More information
‣
Extra sensor heating (Xheat) option (page 16)
4.8 Analog output configuration commands
The analog outputs are set in the factory according to the order form. To change the settings,
use the configuration commands.
You cannot use the serial line commands
AMODE
,
ASEL
and
ATEST
with the
passive output version of HMP155.