When you install HMP155 with the additional T-probe, make sure that the humidity probe and
the T-probe are installed in the same conditions to get accurate readings. Even a slight
difference in temperature conditions result in false RH readings. Install the probes so that the
humidity probe does not heat the T-probe, but is close enough for the probes to share the
same environment conditions. When an RH reading is required, always install the additional T-
probe in the place where you need the reading from.
You can calibrate the additional temperature probe in a liquid bath.
2.5 Connection cover option
Figure 4 HMP155 with optional connection cover
An optional connection cover is available for enhanced protection in wet environments such as
coasts or rainforests.
2.6 Safety
This product has been tested for safety. Note the following precautions:
Ground the product and verify outdoor installation grounding
periodically. Failure to provide proper grounding can result in injury or death
from electrical shock and can severely damage the equipment.
WARNING!
HMP155 User Guide
M210912EN-D
18