HMI41 AND HMP42
M210317EN-A
Operating Manual
39
8.3
Setting the automatic power-off function
set
The number (or text NO) on the first line of the display indicates in minutes
(1...60) the time that the HMI41 stays on before it automatically turns itself
off if no buttons are pressed. The number is changed with buttons
▲
(number
up) and
▼
(number down). If NO is chosen, the automatic power-off function
is not activated. If you do not wish to change any other settings, press the
ON/OFF button. If you wish to change some other settings, press ENTER; the
display changes to show the selection of the display quantities.
8.4
Selecting the display quantities
set
The HMI41 displays relative humidity, temperature and dewpoint temperature
readings. In addition to these, one of the following quantities can be chosen:
absolute humidity, wet bulb temperature and mixing ratio. The number on the
display indicates the following quantities:
0 = RH, T, Td
1 = RH, T, Td, abs
2 = RH, T, Td, Tw
3 = RH, T, Td, x
The number is changed with buttons
▲
(number up) and
▼
(number down). If
you do not wish to change other settings, press the ON/OFF button. If you
wish to change other settings, press ENTER; the display changes to show the
setting of the pressure for mixing ratio and wet bulb temperature calculations.