HMI41 AND HMP42
M210317EN-A
Operating Manual
11
5.
CALIBRATION
5.1
How to use the HMP42 calibration adapter
As the HMP42 probe diameter is extremely small, the probe has a special
calibration adapter for inserting the probe into a salt chamber (e.g. in the
HMK15 calibrator).
NOTE
Be extremely careful when inserting the probe into the
calibration adapter: do not bend or break it.
The probe is inserted into the adapter
with the steel grid on
; this protects the
sensor and prevents it from breaking. Unlike other humidity probes, the
HMP42 is calibrated
with the steel grid on
and thus, the probe naturally
requires a slightly longer stabilization time (approx. 30 minutes).
Figure 5.1.1
HMI41, HMP42 and calibration accessories
Figure 5.1.2
Insert the probe head into the calibration adapter