Value of
Register
0201
hex
Error Message
Severity
1024
Humidity sensor failure (open circuit).
Error
16384
Temperature sensor failure (open circuit).
Error
32768
Temperature sensor failure (short circuit).
Error
262144
Temperature sensor failure (open circuit).
Error
524288
Temperature sensor failure (short circuit).
Error
4194304
Temperature sensor failure (open circuit).
Error
8388608
Temperature sensor failure (short circuit).
Error
Table 17 Error Codes in Register 0203
hex
(32-bit)
Value of
Register
0203
hex
Error Message
Severity
0
Status OK.
256
A/D converter failure (external).
Error
512
Non-volatile memory read/write failure.
Error
4096
Humidity sensor failure (short circuit).
Error
8192
Purge in progress.
Info
16384
A/D converter failure (internal).
Error
32768
Calibration is about to expire.
Info
65536
Calibration has expired.
Warning
A.4.4 Device Identification Objects
Table 18 Device Identification Objects
Object Id
(decimal)
Object Id
(hexadecimal)
Object Name
Example Contents
0
00
hex
VendorName
“Vaisala”
1
01
hex
ProductCode
HPP272
2
02
hex
MajorMinorVersion Software version (for example “1.2.3”)
3
03
hex
VendorUrl
“http://www.vaisala.com/”
HPP272 User Guide
M211972EN-B
56
Содержание HPP272
Страница 1: ...M211972EN B User Guide Vaisala PEROXCAP Hydrogen Peroxide Humidity and Temperature Probe HPP272...
Страница 62: ...Communication Description Value to write 0B40hex 2880 minutes 48 hours HPP272 User Guide M211972EN B 60...
Страница 64: ...HPP272 User Guide M211972EN B 62...
Страница 65: ......
Страница 66: ...www vaisala com...