Only the HUMICAP
â
R2 sensor is designed to be replaced by the
user. Other humidity sensor types available for HMP series probes are composite
sensors where the humidity sensor and temperature sensor are permanently
attached together. If you need to replace a composite sensor, contact a Vaisala
Service Center.
CAUTION!
Sensors are easily damaged when the filter is not in place. Handle
the probe head carefully.
CAUTION!
Reverting the probe to factory settings clears the adjustment values
entered in this procedure. If you need to revert the probe to factory settings
after replacing the sensor yourself, write down the values on the
Adjustment
data
page beforehand, and enter them again after applying the factory settings.
CAUTION!
2
1
Figure 22 HMP3 probe head with filter removed
1
Pt100 temperature sensor
2
HUMICAP
â
R2 humidity sensor
1. Put on clean gloves before touching the filter.
2. Turn the filter counter-clockwise to loosen it.
3. Remove the filter from the probe head. Be careful not to touch the sensors with the filter.
4. There are two sensors under the filter, the HUMICAP
â
sensor and a temperature sensor.
Identify the HUMICAP
â
sensor and make sure it is of the R2 type (temperature sensor not
attached to it).
HMP3, HMP4, HMP5, HMP7, HMP8, HMP9, MMP8, TMP1 User Guide
M212022EN-D
38
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备 气象科学仪器设备
Tel:029-88186182 Web:www.xavery.cn Email:[email protected]
Содержание HMP7
Страница 94: ...陕西威瑞仪器仪表有限公司 生产代理销售 实验检测设备 气象科学仪器设备 Tel 029 88186182 Web www xavery cn Email sxvery 163 com ...
Страница 95: ...陕西威瑞仪器仪表有限公司 生产代理销售 实验检测设备 气象科学仪器设备 Tel 029 88186182 Web www xavery cn Email sxvery 163 com ...
Страница 96: ...www vaisala com 陕西威瑞仪器仪表有限公司 生产代理销售 实验检测设备 气象科学仪器设备 Tel 029 88186182 Web www xavery cn Email sxvery 163 com ...