6.2 Changing the probe filter
• New compatible filter
• Clean lint-free gloves
Sensors are easily damaged when the filter is not in place. Handle
the probe head carefully.
CAUTION!
HMP9 and TMP1 probe models do not have a removable filter.
1. Put on clean gloves before touching the filter.
2. Turn the filter counter-clockwise to loosen it.
3. Remove the filter from the probe head. Be careful not to touch the sensors with the filter.
4. Install a new filter on the probe head. Tighten the filter properly (recommended force
5 Nm).
6.3 Replacing the HUMICAP R2 sensor
• New HUMICAP
â
R2 sensor
• New compatible filter
• Clean lint-free gloves
• 11.3 %RH and 75.5 %RH humidity references (using Vaisala HMK15 Humidity
Calibrator)
• Computer with Microsoft Windows
â
operating system and Vaisala Insight
software installed
• USB connection cable (no. 242659)
Follow this procedure to replace a HUMICAP
â
R2 humidity sensor it has been damaged, or
normal adjustment is not sufficient to restore the humidity measurement accuracy. It is
recommended that you replace the filter at the same time. The procedure includes a special 2-
point adjustment of humidity measurement using Insight PC software. This adjustment must
be done every time after the HUMICAP
â
R2 sensor has been replaced.
Chapter 6 – Maintenance
37
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备 气象科学仪器设备
Tel:029-88186182 Web:www.xavery.cn Email:[email protected]
Содержание HMP7
Страница 94: ...陕西威瑞仪器仪表有限公司 生产代理销售 实验检测设备 气象科学仪器设备 Tel 029 88186182 Web www xavery cn Email sxvery 163 com ...
Страница 95: ...陕西威瑞仪器仪表有限公司 生产代理销售 实验检测设备 气象科学仪器设备 Tel 029 88186182 Web www xavery cn Email sxvery 163 com ...
Страница 96: ...www vaisala com 陕西威瑞仪器仪表有限公司 生产代理销售 实验检测设备 气象科学仪器设备 Tel 029 88186182 Web www xavery cn Email sxvery 163 com ...