HMI41 Calibration Option

19165ZZ-U220en-1.1

Operating Manual

1996-08-29

1

1.

INTRODUCTION

1.1

HMI41 with calibration option

The HMI41 can be used as a field calibrator for various Vaisala humidity
transmitters. There are three different calibration cables available:

 

19116ZZ

for the calibration of Vaisala’s HMD/W 60/70 and 
HMP140 series analogue transmitters

 

19164ZZ

for the calibration of HMP230 series digital transmitters

 

19165ZZ

for the calibration of the HMD/W 20/30 and 
HMP130 series analogue transmitters

In this manual, only the calibration of HMD/W 20/30 and HMP130 series
analogue transmitters is dealt with. For other calibration options, please refer
to corresponding manuals.

Calibration is quick and easy to perform. In calibration, the HMI41 measures
relative humidity with the reference probe and receives the RH information
from the transmitter. These two values are then compared on the HMI41
display and the transmitter is adjusted with calibration potentiometers inside
the transmitter.

1.2

Selecting the HMI41 calibrator function

The desired calibrator function is selected in the HMI41 setup menu. To enter
the setup mode, first press the ON/OFF button and the following appears:

RH

setminmax

hyst hold Lo batHi

P

Pa

h

g/m

T

Td

°C °F

°C °F

%

Then release the ON/OFF button and within 1...2 seconds press both ENTER
and MODE buttons until the following text appears on the display:

After a few seconds, the text changes automatically to show the following:

Содержание HMD20

Страница 1: ...USER S GUIDE Calibration of Series HMD W20 30 and HMP130 Transmitters with Vaisala HUMICAP Indicator HMI41 U220EN11...

Страница 2: ...echanical including photocopying nor may its contents be communicated to a third party without prior written permission of the copyright holder The contents are subject to change without prior notice...

Страница 3: ...20en 1 1 Operating Manual 1996 08 29 i Contents 1 INTRODUCTION 1 1 1 HMI41 with calibration option 1 1 2 Selecting the HMI41 calibrator function 1 2 CALIBRATION 3 2 1 Getting started 3 2 2 Calibration...

Страница 4: ...HMI41 Calibration Option Operating Manual 19165ZZ U220en 1 1 ii 1996 08 29 This page intentionally left blank...

Страница 5: ...libration options please refer to corresponding manuals Calibration is quick and easy to perform In calibration the HMI41 measures relative humidity with the reference probe and receives the RH inform...

Страница 6: ...librator for HMD W 60 70 and HMP140 series analogue transmitters 3 HMI41 as a calibrator for HMP230 series digital transmitters 4 HMI41 as a calibrator for the HMD W20 30 and HMP130 series analogue tr...

Страница 7: ...may vary from 10 minutes to a couple of hours 2 1 Getting started After having selected the desired function Chapter 1 2 turn the HMI41 off and connect the calibration cable to the EXT connector at t...

Страница 8: ...h the ON OFF button and wait until a text similar to the following appears on the display RH Numbers on the first line indicate the RH measured by the transmitter and numbers on the second line indica...

Страница 9: ...ference probe If you wish you can now calibrate the transmitter by pressing MODE again the measured values appear on the display Wait until the readings have stabilized and adjust the calibration pote...

Страница 10: ...unction press ON OFF until you can see some text on the display Then release the ON OFF button and press within 1 2 seconds both ENTER and MODE buttons until the text setup appears on the display Then...

Страница 11: ...ge RH Let the readings stabilize and then adjust the transmitter potentiometers until the readings are the same offset 65 RH gain 65 RH Alternatively you can adjust the transmitter when the display sh...

Страница 12: ......

Страница 13: ......

Страница 14: ...www vaisala com U220EN...

Отзывы: