Chapter 6 _______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 129
9.
Give the
CALIBRATE CHECK
command. The command takes
about 90 seconds to complete.
10. The value produced by the check command must be close to the
value printed on the scattering plates. If the difference is less than
3%, the calibration is correct. In this case, leave the opaque glass
plates in place, write down the signal value, and continue with
section FSM102 Mechanical Alignment Check Procedure on page
130. If the difference is more than 3%, leave the opaque glass
plates in place and continue with the calibration procedure.
0201-077
Figure 41
Assembling and Attaching Calibrator
FSM102 Calibration Procedure
The calibration procedure is performed as follows:
1.
Follow the steps in section FSM102 Calibration Check Procedure
on page 128 (if not yet done).
2.
Install the calibrator assembly on the measurement unit (if not yet
done).